Title Performance evaluation of metal-coated fiber Bragg grating sensors for sensing cryogenic temperature
 
Creator RAJINI-KUMAR, R
SUESSER, M
NARAYANKHEDKAR, KG
KRIEG, G
ATREY, MD
 
Subject temperature sensors
optical techniques
metals
instrumentation
magnetic measurements
 
Description A metal recoated fiber Bragg grating sensor has been designed, fabricated and tested to study its temperature response at cryogenic temperature around 15 K. Metals like aluminium (Al), copper (Cu), lead (Pb) and indium (In) are considered for recoating the FBG sensors. Indium recoated FBG (IRCFBG) sensors showed a higher sensitivity at around 15 K compared to Al, Cu and Pb coated FBG sensors. In addition, the IRCFBG sensor was tested for its repeatability, stability and performance in the high magnetic field (8 T). The experimental results of the IRCFBG sensor test revealed a satisfactory performance. They will be reported in this paper. (C) 2008
 
Publisher ELSEVIER SCI LTD
 
Date 2011-07-22T09:57:31Z
2011-12-26T12:52:22Z
2011-12-27T05:36:38Z
2011-07-22T09:57:31Z
2011-12-26T12:52:22Z
2011-12-27T05:36:38Z
2008
 
Type Article
 
Identifier CRYOGENICS, 48(3-4), 142-147
0011-2275
http://dx.doi.org/10.1016/j.cryogenics.2008.02.007
http://dspace.library.iitb.ac.in/xmlui/handle/10054/6187
http://hdl.handle.net/10054/6187
 
Language en