Impedance Simulator for Testing of Instruments for Bioimpedance Sensing

DSpace at IIT Bombay

View Archive Info
 
 
Field Value
 
Title Impedance Simulator for Testing of Instruments for Bioimpedance Sensing
 
Creator PANDEY, VK
PANDEY, PC
SARVAIYA, JN
 
Subject bioimpedance sensing
testing and calibration
impedance simulator
 
Description Bioimpedance sensing is a noninvasive technique for measuring parameters related to tissue structure or physiological events. Generally, the impedance is sensed by injecting a high frequency low intensity current through a pair of electrodes placed across the selected region of the body and monitoring the voltage developed across the same or another pair of electrodes. The base value of the impedance and its variation can be used, with the help of an appropriate model, for obtaining diagnostic information. For testing and calibration of instruments developed for bioimpedance sensing, we have developed an impedance simulator by using a microcontroller and analog switches. It can be used for measuring sensitivity and frequency response for bioimpedance signals, and for studying the effect of various electrode configurations and common mode interference caused by bioelectric sources and external pickups.
 
Publisher MEDKNOW PUBLICATIONS
 
Date 2011-08-18T17:12:58Z
2011-12-26T12:55:50Z
2011-12-27T05:42:52Z
2011-08-18T17:12:58Z
2011-12-26T12:55:50Z
2011-12-27T05:42:52Z
2008
 
Type Article
 
Identifier IETE JOURNAL OF RESEARCH, 54(3), 203-207
0377-2063
http://dspace.library.iitb.ac.in/xmlui/handle/10054/10079
http://hdl.handle.net/10054/10079
 
Language en