UV photo-ionization based asymmetric field differential ion mobility sensor for trace gas detection
DSpace at IIT Bombay
View Archive InfoField | Value | |
Title |
UV photo-ionization based asymmetric field differential ion mobility sensor for trace gas detection
|
|
Creator |
SURESH, M
VASA, NJ AGARWAL, V CHANDAPILLAI, J |
|
Subject |
Differential ion mobility
Asymmetric wave electric field UV photo-ionization Volatile organic compounds Trace gas sensing FA-IMS MASS-SPECTROMETRY BREATH ANALYSIS EXHALED BREATH EXPOSURE CANCER MS |
|
Description |
A high-field Asymmetric Waveform Ion Mobility Spectrometry (FA-IMS) with ultra-violet (UV) photo-ionization source is proposed and demonstrated for measurement of trace amounts of volatile organic compound (VOC) gases. Experimental studies performed with the sensor using a high frequency asymmetrical waveform differential field for detection of acetone, hexane and acetone-hexane mixture in trace concentrations are discussed. The detection limit as estimated for a signal-to-noise ratio of 3 is of the order of 100 ppb for acetone and hexane. Experimental studies clearly show that selective sensing of a gas in a mixture of gases is also possible by appropriate value for the compensation voltage. Numerical Simulation is also performed using two-dimensional fluid-flow equations to estimate the motion of ions in an electric field. The study also demonstrated influence of parameters such as flow velocity, duty-ratio, on compensation voltage for detection of trace VOC gases with asymmetrical high-frequency electric field. Results based on theoretical study are in agreement with the experimental studies. (C) 2014 Elsevier B.V. All rights reserved.
|
|
Publisher |
ELSEVIER SCIENCE SA
|
|
Date |
2014-12-29T05:15:40Z
2014-12-29T05:15:40Z 2014 |
|
Type |
Article
|
|
Identifier |
SENSORS AND ACTUATORS B-CHEMICAL, 19544-51
0925-4005 http://dx.doi.org/10.1016/j.snb.2014.01.008 http://dspace.library.iitb.ac.in/jspui/handle/100/17169 |
|
Language |
English
|
|