Record Details

Automated derivatization with o-phthalaldehyde for the estimation of amino acids in plasma using reversed-phase high performance liquid chromatography

NOPR - NISCAIR Online Periodicals Repository

View Archive Info
 
 
Field Value
 
Title Automated derivatization with o-phthalaldehyde for the estimation of amino acids in plasma using reversed-phase high performance liquid chromatography
 
Creator Gnanou, J V
Srinivas, S K
Kurpad, A V
 
Subject o-Phthaladehyde derivatization
reversed-phase high performance liquid chromatography
plasma amino acid analysis.
 
Description 322-325
A fully automated method for quantitative estimation of plasma amino acids using fluorescence detection of o-phthaladehyde/2-mercaptoethanol derivatives of the analytes and their separation by gradient elution reversed-phase HPLC has been described. The method is simple and the three-step gradient elution is suitable for routine analysis of a large number of biological samples due to clear resolution, high degree of precision, accuracy, cost-effectiveness and lack of interference from chemical contaminations. Using this method, 19 amino acids were completely resolved and the within-run coefficients of variation ranged from 2.53 to 10.7% with a mean variation of 5.68%.
 
Date 2009-04-02T10:42:21Z
2009-04-02T10:42:21Z
2004-12
 
Type Article
 
Identifier 0301-1208
http://hdl.handle.net/123456789/3725
 
Language en_US
 
Relation G01 N 33
88
 
Publisher CSIR
 
Source IJBB Vol.41(6) [December 2004]