Inter-Laboratory Comparision of NO2 and So2 Generated by Dynamic Dilution System Under Laboratory Contitions:A Technical Discussion.

IR@CSIR-NEERI

View Archive Info
 
 
Field Value
 
Title Inter-Laboratory Comparision of NO2 and So2 Generated by Dynamic Dilution System Under Laboratory Contitions:A Technical Discussion.
 
Creator Khaparde, V V
 
Subject Air Pollution
 
Description A workshop on analytical quality control (AQC) of ambient air quality measurement
methods for nitrogen dioxide (NO2) and sulphur dioxide (SO2) was conducted by Central Pollution
Control Board (CPCB) for officials involved in National Ambient Air Quality Monitoring (NAAQM)
in India. Concentrations of NO2 and SO2 were generated by dynamic dilution system under laboratory
conditions at low and high levels and measured using static dilution system and wet chemical methods
laid down by CPCB under section 16(2)(h) of the air act 1981. CPCB provided the measured values
as reference values for comparing the means obtained by the officials participated from thirteen
organizations. A tolerance limit of ±15% of the reference values was specified to accept the results.
Generated concentrations, which were unknown to the participants, were measured using gaseous
sampling assembly (Envirotech APM 411, New Delhi, India), and wet chemical methods laid down
by CPCB i.e. the same methodology which is used by the organizations to generate the data of NO2 and
SO2 in ambient air. Simultaneously, concentrations were checked by CPCB using automatic analyzers
as a check on reference concentration. It is observed that results of automatic analyzers for NO2 and
SO2 were within a tolerance of ±5% with %RSD below 3. On the other hand, results of most of the
participants showed variability in the measurements with %RSD ranging between ±0.8 and ±88.6
and exceedences of means from the tolerance limit with bias ranging between 1.4 and −59%. To
check the cause of high variability in the measurements obtained under identical conditions, duplicate
sampling was performed by one of the participants for SO2 at low concentration level. In this study,
results of wet chemical methods, automatic analyzers and results of duplicate sampling are analysed
statistically to assess the cause of high variability in the measurements. Analysis of t-test and analysis
of variance (ANOVA) showed highly significant results for NO2 and SO2 at high concentration levels
(α 0.05) and for SO2 at both the levels (α 0.01) respectively indicating some bias is existing either in
the sampling or in analytical technique. Duplicate sampling performed to check precision in parallel
measurements showed high %RSD indicating the presence of systematic error in sampling technique
as the same calibration factor (CF) was used to measure the concentration of duplicate samples.
Statistical analysis of flow rates of duplicate sampling showed that the sampling assembly could not
maintain the constant flow rate within the ±10% with that measured at the start of the sampling.
This resulted in high %RSD and deviation from the reference values for the results of most of the
participants, even after accepting ±15% tolerance limit. There is a need to improve and evaluate this
gaseous sample collection device under laboratory conditions to generate reliable database of NO2
and SO2 in ambient air.
 
Publisher Springer
 
Date 2006-05
 
Type Article
PeerReviewed
 
Format application/pdf
 
Identifier http://neeri.csircentral.net/46/1/Khaparde1.pdf
Khaparde, V V (2006) Inter-Laboratory Comparision of NO2 and So2 Generated by Dynamic Dilution System Under Laboratory Contitions:A Technical Discussion. Environmental Monitoring and Assessment, 116 (1). pp. 217-231. ISSN 0167-6369 (Print) 1573-2959 (Online)
 
Relation https://link.springer.com/journal/10661