<strong>Comparison of measured and Monte Carlo-calculated electron depth dose distributions in aluminium</strong>

Online Publishing @ NISCAIR

View Archive Info
 
 
Field Value
 
Authentication Code dc
 
Title Statement <strong>Comparison of measured and Monte Carlo-calculated electron depth dose distributions in aluminium</strong>
 
Added Entry - Uncontrolled Name Shrivastava, Vandana ; Bhabha Atomic Research Centre
Benny, P. G. ; Bhabha Atomic Research Centre
Palani Selvam, T. ; Bhabha Atomic Research Centre
 
Uncontrolled Index Term radiation dosimetry
Electron parameters; Aluminium; Electron beam accelerator; Dosimetry
 
Summary, etc. Depth dose profiles in aluminium have been measured using the cellulose triacetate dosimeter against different electron energies (4, 4.5 and 5 MeV) at a recently upgraded 15 kW industrial electron beam accelerator facility. The study also includes comparison of these profiles against Monte Carlo calculations. The measured and simulated depth dose profiles are similar in shape. For all electron energies, at initial depths, the measured doses are higher than the simulated ones. The simulated and measured normalized surface dose values are 0.58 and 0.66, respectively, independent of electron energy. The difference in the surface dose between Monte Carlo and experiment could be attributed to possible presence of low energy electrons in the measurements whereas the Monte Carlo calculations are based on monoenergetic electrons. Between the region of dose maximum and the tail portion of the depth dose curve, the measured dose is smaller than the simulated values (about 17% to 40% at 5 MeV). Using the depth dose profiles, electron beam parameters such as depth at which maximum dose occurs, <em>d</em><sub>max</sub>, practical range, <em>R</em><sub>p</sub> and half-value depth, <em>R</em><sub>50</sub> have been determined. Using the measured parameters <em>R</em><sub>p </sub> and <em>R</em><sub>50</sub>, the incident kinetic energy of the electron beam has been determined. The estimated electron energies while using <em>R</em><sub>p</sub> are 4.02, 4.41 and 4.75 MeV. When using <em>R</em><sub>50</sub>, the corresponding values are 3.83, 4.21 and 4.64 MeV. The measured <em>R</em><sub>P</sub>/<em>R</em><sub>50 </sub> ratios are slightly larger than the Monte Carlo-calculated values, which suggest that the electron beam may not be monoenergetic.
 
Publication, Distribution, Etc. Indian Journal of Pure & Applied Physics (IJPAP)
2018-01-29 09:25:55
 
Electronic Location and Access application/pdf
http://op.niscair.res.in/index.php/IJPAP/article/view/16660
 
Data Source Entry Indian Journal of Pure & Applied Physics (IJPAP); ##issue.vol## 56, ##issue.no## 01 (2018): Indian Journal of Pure & Applied Physics
 
Language Note en
 
Nonspecific Relationship Entry http://op.niscair.res.in/index.php/IJPAP/article/download/16660/49870
 
Terms Governing Use and Reproduction Note Except where otherwise noted, the Articles on this site are licensed under Creative Commons License: CC Attribution-Noncommercial-No Derivative Works 2.5 India © 2015. The Council of Scientific &amp; Industrial Research, New Delhi.