Sensitivity Enhancement of Electrochemical Biosensor for Point of Care (POC) Applications: Vi Antigen Detection as a Case Study
DIR@IMTECH: CSIR-Institute of Microbial Technology
View Archive InfoField | Value | |
Title |
Sensitivity Enhancement of Electrochemical Biosensor for Point of Care (POC) Applications: Vi Antigen Detection as a Case Study
|
|
Creator |
Maurya, Ranjan Kumar
Alam, M. A. Ahamad, Nadeem Kishore, Kaushal Prajesh, Rahul Choudhary, Megha Bhalla, Vijayender Agarwal, Ajay |
|
Subject |
QR Microbiology
|
|
Description |
A novel approach for signal enhancement of electrochemical biosensors by incorporating mechanical vibrations has been developed. We report the electrochemical study of the ferricyanide and ferrocyanide $\left(\,{\left[{\rm{Fe}}{\left({\rm{CN}}\right)}_{6}\right]}^{-3/-4}\right)$ a redox couple, at room temperature (∼25 °C), under the effect of mechanical vibrations of different frequencies applied to the sensor, for sensitivity enhancement. The experimental results showed a sensitivity enhancement of ∼332% (from $3.125\,{\rm{nA}}\,\mu {{\rm{M}}}^{-1}$ to $13.5\,{\rm{nA}}\,\mu {{\rm{M}}}^{-1}$) at 88.75 Hz of vibration. This novel approach of signal sensitivity enhancement is also validated with antibody immobilization-based Vi antigen detection for typhoid assessment. The sensitivity enhancement up to ∼15% is achieved for Vi antigen detection under the mechanical vibration of 120 Hz. The sensor is fabricated using microfabrication technology. The vibration-assisted ultrasensitive biosensing platform is also prototyped as a portable and Internet of Thing (IoT) enabled device, suitable for point-of-care applications. Detailed features of the prototype along with the test results are elaborated in the paper
|
|
Publisher |
IOP SCIENCE
|
|
Date |
2021-01-14
|
|
Type |
Article
PeerReviewed |
|
Relation |
https://iopscience.iop.org/article/10.1149/1945-7111/abd929
http://crdd.osdd.net/open/2777/ |
|
Identifier |
Maurya, Ranjan Kumar and Alam, M. A. and Ahamad, Nadeem and Kishore, Kaushal and Prajesh, Rahul and Choudhary, Megha and Bhalla, Vijayender and Agarwal, Ajay (2021) Sensitivity Enhancement of Electrochemical Biosensor for Point of Care (POC) Applications: Vi Antigen Detection as a Case Study. JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 168 (1).
|
|