<strong>Gamma Passing Rate – Correlation with Patient Specific Quality Assurance Devices using Two Photon energies in Pelvic IMRT</strong>
Online Publishing @ NISCAIR
View Archive InfoField | Value | |
Authentication Code |
dc |
|
Title Statement |
<strong>Gamma Passing Rate – Correlation with Patient Specific Quality Assurance Devices using Two Photon energies in Pelvic IMRT</strong> |
|
Added Entry - Uncontrolled Name |
NIGAM, JITENDRA ; Department of Physics
Invertis University
Bareilly 243 123 KUMAR, PIYUSH ; SHRI RAM MURTI SMARAK INSTITUTE OF MEDICAL SCIENCES, BAREILLY 243202 SINGH, P P; DEPARTMENT OF PHYSICS Invertis University Bareilly-243 123 S, SILAMBARASAN N; SHRI RAM MURTI SMARAK INSTITUTE OF MEDICAL SCIENCES, BAREILLY 243202 S, NAVITHA ; SHRI RAM MURTI SMARAK INSTITUTE OF MEDICAL SCIENCES, BAREILLY 243202 |
|
Uncontrolled Index Term |
MEDICAL PHYSICS; RADIATION PHYSICS; INTERDISCIPLINARY Gamma Analysis; IMRT; MatriXX; EPID; PSQA; Pelvis; Gamma Passing Rate; 6 MV & 10 MV Photon Energy |
|
Summary, etc. |
The effect on Gamma Passing Rate is studied by changing the Photon Energy in the Patient Specific Quality Assurancein Pelvic IMRT using EPID and MatriXX keeping PSQA device at Isocenter. Dose verification of 62 patients, sufferingfrom pelvic malignancies and treated with Intensity modulated radiation therapy (IMRT), with EPID and MatriXX usinggamma criteria of 3%/3 mm, 3%/2 mm, 2%/3 mm, 2%/2 mm and investigated the effect on Gamma Passing Rate (%GP) bychanging beam energy from 6 MV to 10 MV. The results demonstrated that the behavior of gamma passing rate decreases ascriteria is changed from 3%/3 mm to strict criteria. The mean gamma passing rate for 6 MV beam were (%GP±StandardDeviation σ) 99.64±0.81, 99.45±1.08, 97.57±2.12, 96.17±2.63 using the criteria 3%/3 mm, 3%/2 mm, 2%/3 mm, 2%/2 mmrespectively in EPID. For MatriXX, the mean %GP were 96.69±4.11, 93.85±5.31, 88.17±10.63, 82.2±12.28 respectively.For 10 MV, the mean (%GP±Std. Dev) were 99.64±0.81, 99.45±1.08, 97.57±2.12, 96.17±2.63 for in EPID, 96.69±4.11,93.85±5.31, 88.17±10.63, 82.2±12.28 for MatriXX respectively. Difference between %GP for 6 MV and 10 MV beam were-0.01, -0.11, 1.08, 1.61 with gamma criteria of 3%/3 mm, 3%/2 mm, 2%/3 mm, 2%/2 in EPID. For MatriXX, differencebetween %GP for 6 MV and 10 MV beam were -0.40, -0.24, 0.96, 1.62 respectively. It can be concluded that applying morestrict gamma criteria results in low gamma passing rate. There is marked difference in Gamma Passing Rate (%GP) withchange in photon energy. |
|
Publication, Distribution, Etc. |
Indian Journal of Pure & Applied Physics (IJPAP) 2022-09-06 11:26:25 |
|
Electronic Location and Access |
application/pdf http://op.niscair.res.in/index.php/IJPAP/article/view/55453 |
|
Data Source Entry |
Indian Journal of Pure & Applied Physics (IJPAP); ##issue.vol## 60, ##issue.no## 9 (2022): Indian Journal of Pure & Applied Physics |
|
Language Note |
en |
|
Nonspecific Relationship Entry |
http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580361 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580362 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580363 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580364 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580365 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580366 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580368 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580369 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580370 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580371 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580372 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580373 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580374 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465580375 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465581354 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465581355 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465581356 http://op.niscair.res.in/index.php/IJPAP/article/download/55453/465581357 |
|
Terms Governing Use and Reproduction Note |
Except where otherwise noted, the Articles on this site are licensed under Creative Commons License: CC Attribution-Noncommercial-No Derivative Works 2.5 India © 2015. The Council of Scientific & Industrial Research, New Delhi. |
|